Vyaire Revel Ventilator

The Vyaire ReVel Portable Critical Care Ventilator is a popular legacy ventilator offering high-performance ventilation for adults and pediatric patients (above 5kg.) The ReVel Ventilator features the patented ActivCore gas delivery system which helps ensure safe transport even for high-acuity patients.

Keep your Vyaire Revel operating safely and reliably with functions tests and repairs. Certified biomedical technicians follow OEM-compliant service protocols to calibrate, test, and verify your equipment meets OEM specs for safety and performance. Need parts, disposables, or accessories? Visit our trusted parts and accessories partner for compatible components.

Service Options

  • PM Cycle: 3 Year
  • PM Cycle: 6 Year
  • Functions Test
  • Repair
  • End-of-Life

Possible Repair or Service Needs

Dirty or clogged filter
Inaccurate flow rate
Clogged blower
Inaccurate readings
Hardware Fault 20 and/or 21 – This issue is an internal PCB issue and requires the unit be sent in for replacement of either Hybrid PCB, Main System PCB, or both
Schd Srvc # (90-99) – may be one or multiple, but is another internal issue dealing with the PCBs and must be sent in
Damaged power pigtail
Not delivering set volume – this could be a few different modules inside of the unit from pressure transducer module, to blower module, to and issue in the exhalation module, all issues require service at the depot
Blower demand – This is normally seen when using the vent in CPAP or NPPV mode, caused by a leak in the setup. This has been generally seen when the patient has a poor seal around the mask being used.

Care Instructions

Proper storage in climate climate-controlled area

Putting a case on the unit

Use only fingers to press buttons (not ink pens)

Be gentle with older units