Defibtech Lifeline ECG AED

The Lifeline VIEW family of defibrillators contains the first and only AEDs with full-motion, high-resolution color video. The Lifeline ECG offers users the ability to view step-by-step instructional videos or ECG information. Additionally, the Lifeline ECG can be used in a non-diagnostic ECG monitoring mode, to allow for rhythm and heart rate monitoring.Keep your Defibtech Lifeline ECG operating safely and reliably with annual Preventative Maintenance (PM) service OR a multi-year service contract. Certified biomedical technicians follow OEM-compliant service protocols to calibrate, test, and verify your equipment meets OEM specs for safety and performance.Need parts, disposables, or accessories? Visit our trusted parts and accessories partner for compatible components.

Lifeline ECG Service Options

  • PM Cycle: Annual
  • Functions Test
  • Preventative Maintenance (PM)Learn More
  • On-Site PMs
  • Service ContractLearn More

Possible Repair or Service Needs for the Lifeline ECG

Issues with charging
Issues with powering on
Failed self test
Pads not detected or registering
Issues with delivering charge (shocking low or not shocking)
Beeping
Shock button not flashing

Preventative Maintenance for the Lifeline ECG

  • Visual inspection to verify integrity of unit

  • Diagnostic check of device’s CPU

  • Testing of power supply system and internally housed battery

  • Testing of all device parameters to OEM specifications

  • Calibration of device functions to OEM specifications

  • Complimentary device cleaning

  • Verify the unit calls for pads to be installed

  • Verify the unit passes self-testing

  • Verify the unit displays a ready-for-use indicator after self-testing

  • Verify the unit runs through the rescue procedure

  • Application of inspection sticker

  • Free return shipping

  • Full biomedical report for compliance and documentation

Service Contract for the Defibtech Lifeline ECG

Everything in Annual PM, AND:

  • Parts and Labor for all repairs and maintenance necessary to meet OEM requirements

  • Asset management via RAMS

  • Automated, scheduled service

  • Access to replacement parts

Lifeline ECG Care Instructions

Proper storage in climate climate-controlled area

Putting a case on the unit

Use only fingers to press buttons (not ink pens)