- Monopolar/Bipolar port not functioning – As the ports are repeatedly used the interal contacts can become loose and need to be retightened
- Does not recognize REM – The pressure sensor that the ren cable plugs into may be old and no longer closing the circuit when plugged in
- Error code 16 – May be corrected with a calibration, if not mroe than likely looking at a bad Control PCB
- Unit does not complete startup when powered on (bootloop) – This has been seen to stem from a DC voltage issue coming from the low voltage power supply or Control PCB
- Error code 17 – Unit is detecting an output without footswitch use, may be an issue witht he RF PCB needing to be replaced
- Damaged nosecone or damages case
The Valleylab Force FX™ generator adjusts automatically, responding to tissue changes, maintaining power delivery, and minimizing drag.