VALLEYLAB FORCE FX/FXC

VALLEYLAB FORCE FX/FXC
SERVICE

REPAIR

PREVENTATIVE MAINTENANCE

FUNCTIONS TEST

Common Issues:

  • Monopolar/Bipolar port not functioning – As the ports are repeatedly used the interal contacts can become loose and need to be retightened
  • Does not recognize REM – The pressure sensor that the ren cable plugs into may be old and no longer closing the circuit when plugged in
  • Error code 16 – May be corrected with a calibration, if not mroe than likely looking at a bad Control PCB
  • Unit does not complete startup when powered on (bootloop) – This has been seen to stem from a DC voltage issue coming from the low voltage power supply or Control PCB
  • Error code 17 – Unit is detecting an output without footswitch use, may be an issue witht he RF PCB needing to be replaced
  • Damaged nosecone or damages case
Description:
The Valleylab Force FX™ generator adjusts automatically, responding to tissue changes, maintaining power delivery, and minimizing drag.